Development of a third-generation electrochemical enzyme-based biosensor for a scalable detection of oxygen in power generation cells
Date
2019-12
Authors
Jiyane, Sphumelele Nomnontho
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Pencil graphite electrodes (PGEs) are another form of carbon electrodes with good
mechanical strength and comparable electrical properties. Moreover. their low cost
makes them an excellent alternative to more conventional electrodes. especially in
disposable applications. In this study. the PGEs were constructed with a 2 mm
diameter pencil graphite with hardness 4H. HB. and 4B. The electrodes were cleaned
and modified with 1 mg/ml of graphene oxide (GO) to enhance the surface area of the
electrodes. The PGE-GO was further reduced electrochemically using Na2S2O4 from
-1.2 V to 0.8 V at 50 mV/s for 50 cyclic voltammetry scans in the presence of oxygen.
using K3Fe(CN)6 / K4Fe(CN)6 as a redox couple. The performance of the PGE was
evaluated with multi-walled carbon nanotubes and nanomaterials with various linking
agents. A further evaluation was conducted with multi-copper oxidase (MCO)
enzymes (Bilirubin oxidase (BOx) and Laccase oxidase) applied for the bio-catalytic
reduction of oxygen.
The outcome of this study showed that the modification with GO revealed redox peaks
3.6 times higher than the bare PGE. The immobilization of MCO was confirmed by
cyclic voltammetry in the presence of a phosphate buffer. Furthermore. the
amperometric measurements of O2 at a reducing potential of +0.34 V. showed linearity
up to 0.36 mM and sensitivity of 520 μA/(mM.cm2) to O2. Furthermore. computational adsorption studies were performed for the layer-by-layer
electrode modification steps. The adsorption simulations revealed a lowering of the
energy favored between the designed electrode layers. suggesting a most favorable
interaction for the GO/MWCNTs/PBSE/BOx layer. Overall the computational data
correlated well with experimental work. Notably. the layer-by-layer adsorption of the
GO/MWCNTs/PBSE/BOx showed excellent affinity 11.4 M−1 between PBSE and the
enzyme interaction. The direct electron transfer (DET) of the enzymatic reaction
integrated with nanotechnology. has led to a small. portable and renewable power
generating device. Thus this study addresses the demand for implantable medical
devices. in the absence of an external power source.
Description
Submitted in fulfilment of the requirements for the Degree of Master of Applied Sciences in Chemistry in the Faculty of Applied Sciences at the Durban University of Technology, 2019.